A miniature intraventricular axial pump for left ventricular (LV) supp
ort is under development. This pump was designed for placement in the
LV cavity by insertion through the LV apex with the outlet located at
the ascending aorta via the aortic valve, The basic hydrodynamic chara
cteristics represented as a relationship between pump head (H) and flo
w (Q) showed a negative linear relationship under a constant head. Thi
s characteristic was generally the same as that obtained by other axia
l rotation pumps. However, the actual H-Q relationship was represented
as anticlockwise ''loops'' caused by the contraction of the natural L
V. The comparative in vitro data on these H-Q loops showed that the sh
ape of the loops was changed drastically by the connecting condition b
etween the pump and natural cardiovascular system.