METERED-DOSE INHALERS .1. DRUG CONTENT AND PARTICLE-SIZE DISTRIBUTIONOF BECLOMETHASONE DIPROPIONATE

Citation
M. Lebelle et al., METERED-DOSE INHALERS .1. DRUG CONTENT AND PARTICLE-SIZE DISTRIBUTIONOF BECLOMETHASONE DIPROPIONATE, Journal of pharmaceutical and biomedical analysis, 14(7), 1996, pp. 793-800
Citations number
19
Categorie Soggetti
Pharmacology & Pharmacy
ISSN journal
07317085
Volume
14
Issue
7
Year of publication
1996
Pages
793 - 800
Database
ISI
SICI code
0731-7085(1996)14:7<793:MI.DCA>2.0.ZU;2-M
Abstract
Four commercially available beclomethasone metered dose inhalers were analyzed for both spray content uniformity and par-tide size. The drug contents of primed and unprimed sprays collected at the beginning of the lifetime of the canister were not significantly different from tho se collected throughout the experiment. Particle size analysis of the four products using the Andersen Cascade Impactor Mark II showed that the distribution profiles were not identical. An existing HPLC method was modified to quantitate single sprays for content uniformity and to measure the amount on an impactor stage for particle sizing.