A CALIBRATION METHOD FOR WHOLE-BODY COUNTERS, USING MONTE-CARLO SIMULATION

Citation
T. Ishikawa et al., A CALIBRATION METHOD FOR WHOLE-BODY COUNTERS, USING MONTE-CARLO SIMULATION, Radiation protection dosimetry, 64(4), 1996, pp. 283-288
Citations number
11
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
64
Issue
4
Year of publication
1996
Pages
283 - 288
Database
ISI
SICI code
0144-8420(1996)64:4<283:ACMFWC>2.0.ZU;2-H
Abstract
A Monte Carlo simulation code was developed to estimate the counting e fficiencies in whole-body counting for various body sizes. The code co nsists of mathematical models and parameters which are categorised int o three groups: a geometrical model for phantoms and detectors, a phot on transport model, and a detection system model. Photon histories wer e simulated with these models. The counting efficiencies for five Cs-1 37 block phantoms of different sizes were calculated by the code and c ompared with those measured with a whole-body counter at NIRS (Japan). The phantoms corresponded to a newborn, a 5 month old, a 6 year old, an 11 year old and an adult. The differences between the measured and calculated values were within 6%. For the adult phantom, the differenc e was 0.5%. The results suggest that the Monte Carlo simulation code c an be used to estimate the counting efficiencies for various body size s.