IMPLEMENTATION OF INTERNAL LABORATORY QUALITY-CONTROL PROCEDURES FOR THE MONITORING OF ELISA PERFORMANCE AT A REGIONAL VETERINARY LABORATORY

Citation
Sd. Blacksell et al., IMPLEMENTATION OF INTERNAL LABORATORY QUALITY-CONTROL PROCEDURES FOR THE MONITORING OF ELISA PERFORMANCE AT A REGIONAL VETERINARY LABORATORY, Veterinary microbiology, 51(1-2), 1996, pp. 1-9
Citations number
10
Categorie Soggetti
Microbiology,"Veterinary Sciences
Journal title
ISSN journal
03781135
Volume
51
Issue
1-2
Year of publication
1996
Pages
1 - 9
Database
ISI
SICI code
0378-1135(1996)51:1-2<1:IOILQP>2.0.ZU;2-X
Abstract
Quality control (QC) procedures for antigen detection enzyme-linked im munosorbent assays (ELISAs) for hog cholera (HC) virus, foot and mouth disease (FMD) virus, and an antibody detection ELISA. for FMD virus w ere established at a regional veterinary laboratory in northern Thaila nd. A recently developed computer software package, QCEL, was used to facilitate management and analysis of QC data. The program was used to assess test performance by producing Shewhart-CUSUM control charts wh ich monitored control data for unacceptable fluctuations or trends. QC EL-generated control charts and analyses are presented and discussed. The use of a simple;integrated computerised system for storage and ana lysis of QC control data provided the laboratory with the opportunity to achieve increased confidence in the results of tests performed.