Gk. Webster et al., OPTIMIZATION AND AUTOMATION OF AOAC OFFICIAL METHOD 971.47 FOR DETERMINATION OF ROXARSONE IN FEED, Journal of AOAC International, 79(4), 1996, pp. 1012-1017
AOAC Official Method 971.47 was optimized for use with solid-phase ext
raction (SPE) using the same activated carbon referenced in the origin
al method, The optimization of AOAC 971.47 begins after the roxarsone
is extracted into the 2% K2HPO4 solution of the original method; inste
ad of 30 mL, 9 mL is pipetted into a test tube, Next, to flocculate th
e protein, 300 mu L HCl is pipetted info the tube and mixed, After 15
min, the contents of the tube are filtered through a 0.45 mu m nylon m
embrane filter equipped with a glass prefilter, Next, 7 mL clear filtr
ate is pipetted into a new test tube, followed, by the addition of 234
mu L NaOH, and the contents of the tube are mixed, The filtrate is th
en passed through a 400 mg bed of activated carbon packed in a 3 mL SP
E tube at a flow rate of 1 mL/min. The spectrophotometric analysis of
the e(sic)uate is not changed from the original AOAC 971.47, The optim
ized method was validated and semiautomated, Results of roxarsone dete
rmination by the optimized procedure were found to be acceptable at al
l levels of approved claims, For the commercial feed samples analyzed,
recoveries by the automated method ranged from 93.1 to 104% of the re
sults reported by an independent laboratory using AOAC 971.47.