Neutron fields in six locations inside containment of a CANDU(R)600-PH
WR power plant were characterized using Bonner-sphere spectrometry, Un
folded fluence spectra were used to predict and understand the behavio
r of a rem meter (a moderator-type dose equivalent survey instrument),
The suitability of employing commonly-used sources such as Am-241-Be
for calibrating the rem meter was investigated by calculational means,
Results of these calculations suggest that employing a calibration fi
eld more representative of the power-plant fields would likely provide
more accurate dose equivalents.