QUALITY-CONTROL OF PHARMACEUTICALS CONTAINING CLENBUTEROL BY THERMAL LENS SPECTROMETRY

Citation
Ym. Biosca et al., QUALITY-CONTROL OF PHARMACEUTICALS CONTAINING CLENBUTEROL BY THERMAL LENS SPECTROMETRY, Journal of pharmaceutical and biomedical analysis, 14(8-10), 1996, pp. 1037-1041
Citations number
19
Categorie Soggetti
Pharmacology & Pharmacy
ISSN journal
07317085
Volume
14
Issue
8-10
Year of publication
1996
Pages
1037 - 1041
Database
ISI
SICI code
0731-7085(1996)14:8-10<1037:QOPCCB>2.0.ZU;2-S
Abstract
An ultrasensitive absorptiometric procedure for the determination of c lenbuterol in pharmaceutical preparations was developed. Clenbuterol w as diazotized with nitrite and coupled with 1-(naphthyl)ethylenediamin e, and the absorbance of the azo dye formed was measured by both spect rophotometry and ultrasensitive thermal lens spectrometry (TLS). The T LS limit of detection was 1.5 ppb, 14-fold lower than with a Hewlett-P ackard diode array spectrophotometer. Thus, the TLS procedure can be a dvantageously applied to quality control of clenbuterol at the individ ual dose level and in small samples. Repeatability as relative standar d deviation was 1.5% (50 ppb, n = 6).