RESOLUTION IMPROVEMENT BY USE OF CARBOXYMETHYL-BETA-CYCLODEXTRIN AS CHIRAL ADDITIVE FOR THE ENANTIOMERIC SEPARATION OF BASIC DRUGS BY CAPILLARY ELECTROPHORESIS
M. Fillet et al., RESOLUTION IMPROVEMENT BY USE OF CARBOXYMETHYL-BETA-CYCLODEXTRIN AS CHIRAL ADDITIVE FOR THE ENANTIOMERIC SEPARATION OF BASIC DRUGS BY CAPILLARY ELECTROPHORESIS, Journal of pharmaceutical and biomedical analysis, 14(8-10), 1996, pp. 1107-1114
Three beta-cyclodextrin derivatives-carboxymethyl-, dimethyl- and hydr
oxypropyl-beta-cyclodextrin-were tested as chiral selectors for the en
antioseparation of seven basic drugs in free solution capillary electr
ophoresis, using buffers made of 100 mM phosphoric acid adjusted to pH
3.0 with triethanolamine in fused silica capillaries thermostatted at
15 degrees C. The best results with respect to chiral resolution were
obtained with carboxymethyl-beta-cyclodextrin (CMCD): the enantiomers
of all compounds examined were completely resolved with this beta-cyc
lodextrin derivative. The influence of the CMCD concentration on the m
igration times, the apparent electrophoretic mobility difference and t
he resolution of the drug enantiomers was investigated thoroughly. Par
ticularly impressive resolution values, up to 23.7, were obtained for
several compounds in these capillary electrophoretic systems, using CM
CD in the 5-15 mM concentration range.