ELECTRONIC IMAGING-SYSTEM USING A HIGH-PERFORMANCE HARPICON TUBE FOR NEUTRON RADIOGRAPHY

Citation
K. Mochiki et al., ELECTRONIC IMAGING-SYSTEM USING A HIGH-PERFORMANCE HARPICON TUBE FOR NEUTRON RADIOGRAPHY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 377(1), 1996, pp. 126-129
Citations number
6
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
377
Issue
1
Year of publication
1996
Pages
126 - 129
Database
ISI
SICI code
0168-9002(1996)377:1<126:EIUAHH>2.0.ZU;2-Z
Abstract
A high-performance HARPICON tube has been equipped to an electronic im aging camera of Musashi Image Processing System for neutron radiograph y (NRG). A vertical limiting resolution of 1600 TV lines was obtained with an imaging size of 2048 x 2048 pixels using the EIAJ rest chart-A . The camera was applied to neutron radiography at the NRG facility of JRR-3M at Japan Atomic Energy Research Institute and excellent neutro n radiographs have been obtained. Durability of the HARPICON camera ag ainst neutron and gamma rays was proved.