OPTIMIZATION OF DETECTION OF ATOMIC-ABSORPTION LINES IN INTRACAVITY LASER SPECTROSCOPY

Citation
Vs. Burakov et al., OPTIMIZATION OF DETECTION OF ATOMIC-ABSORPTION LINES IN INTRACAVITY LASER SPECTROSCOPY, Fresenius' journal of analytical chemistry, 355(7-8), 1996, pp. 883-886
Citations number
11
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
355
Issue
7-8
Year of publication
1996
Pages
883 - 886
Database
ISI
SICI code
0937-0633(1996)355:7-8<883:OODOAL>2.0.ZU;2-D
Abstract
An intracavity laser spectrometer equipped with a graphite furnace ele ctrothermal atomizer and two alternative types of narrow atomic lines detection schemes (high resolution diffraction spectrograph with optic al multichannel analyzer or a resonant detector based on a hollow-cath ode lamp) is described. Such system was used to determine ultra-trace amounts of lithium and strontium in aqueous solutions. A significant r eduction in the measurable absorbance was demonstrated for both elemen ts. Careful optimization of the operating conditions of the detection systems and a comparison of their typical features and advantages were performed.