COMPARISON OF CONVENTIONAL SINGLE ECHO AND MULTI-ECHO SEQUENCES WITH A FAST SPIN-ECHO SEQUENCE FOR QUANTITATIVE T2 MAPPING - APPLICATION TOTHE PROSTATE

Citation
Gp. Liney et al., COMPARISON OF CONVENTIONAL SINGLE ECHO AND MULTI-ECHO SEQUENCES WITH A FAST SPIN-ECHO SEQUENCE FOR QUANTITATIVE T2 MAPPING - APPLICATION TOTHE PROSTATE, Journal of magnetic resonance imaging, 6(4), 1996, pp. 603-607
Citations number
14
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
10531807
Volume
6
Issue
4
Year of publication
1996
Pages
603 - 607
Database
ISI
SICI code
1053-1807(1996)6:4<603:COCSEA>2.0.ZU;2-8
Abstract
The accuracy of water T2 maps generated from a fast spin-echo (FSE) se quence was compared with data obtained by conventional single and mult i-echo spin-echo pulse sequences using a commercial gel phantom, Spati ally localized stimulated echo acquisition mode (STEAM) proton spectro scopy was also used to confirm the reported water T2 values of the gel s contained in the phantom, The FSE sequence was shown to be superior in accuracy to both the single and multi-echo spin echo sequences and comparable to STEAM, producing results that were within 10% of known v alues. The effectiveness of the FSE sequence was further demonstrated by generating T2 maps of the normal and diseased prostate in clinicall y acceptable imaging times, resulting in comparable T2 values to those obtained using STEAM. Accurate quantitative T2 maps can be produced w ith the FSE sequence.