COMPARISON OF CONVENTIONAL SINGLE ECHO AND MULTI-ECHO SEQUENCES WITH A FAST SPIN-ECHO SEQUENCE FOR QUANTITATIVE T2 MAPPING - APPLICATION TOTHE PROSTATE
Gp. Liney et al., COMPARISON OF CONVENTIONAL SINGLE ECHO AND MULTI-ECHO SEQUENCES WITH A FAST SPIN-ECHO SEQUENCE FOR QUANTITATIVE T2 MAPPING - APPLICATION TOTHE PROSTATE, Journal of magnetic resonance imaging, 6(4), 1996, pp. 603-607
The accuracy of water T2 maps generated from a fast spin-echo (FSE) se
quence was compared with data obtained by conventional single and mult
i-echo spin-echo pulse sequences using a commercial gel phantom, Spati
ally localized stimulated echo acquisition mode (STEAM) proton spectro
scopy was also used to confirm the reported water T2 values of the gel
s contained in the phantom, The FSE sequence was shown to be superior
in accuracy to both the single and multi-echo spin echo sequences and
comparable to STEAM, producing results that were within 10% of known v
alues. The effectiveness of the FSE sequence was further demonstrated
by generating T2 maps of the normal and diseased prostate in clinicall
y acceptable imaging times, resulting in comparable T2 values to those
obtained using STEAM. Accurate quantitative T2 maps can be produced w
ith the FSE sequence.