PERFORMANCE OF A GE-BGO COMPTON-SUPPRESSION SPECTROMETER AND ITS APPLICATION TO PHOTON-ACTIVATION ANALYSIS

Citation
K. Fukuda et al., PERFORMANCE OF A GE-BGO COMPTON-SUPPRESSION SPECTROMETER AND ITS APPLICATION TO PHOTON-ACTIVATION ANALYSIS, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 114(3-4), 1996, pp. 379-386
Citations number
12
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
114
Issue
3-4
Year of publication
1996
Pages
379 - 386
Database
ISI
SICI code
0168-583X(1996)114:3-4<379:POAGCS>2.0.ZU;2-C
Abstract
A Compton-suppression spectrometer, consisting of a pure Ge detector a nd EGO shield detectors is described. A cylindrical symmetry type of t he shield detector configuration, with the Ge detector inserted on the symmetry axis, has been designed. The Compton-suppression factor is 6 .0 for the photopeak of Co-60 at 1332.5 keV. The peak-to-Compton ratio for the peak is 42:1 when the BGO detectors are used as a passive shi eld only, and is 250:1 in the suppression mode. Photon activation anal ysis with this detector has been made for the determination of trace e lements in a bulk dysprosium sample. The signal-to-noise ratio for det ecting trace elements in the sample has been improved by a factor from 1.3 to 3.1 by Compton suppression compared with that obtained in the non-suppression mode.