TESTS OF A 3RD-GENERATION MULTIANODE PHOTOTUBE

Citation
Y. Bonushkin et al., TESTS OF A 3RD-GENERATION MULTIANODE PHOTOTUBE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 381(2-3), 1996, pp. 349-354
Citations number
11
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
381
Issue
2-3
Year of publication
1996
Pages
349 - 354
Database
ISI
SICI code
0168-9002(1996)381:2-3<349:TOA3MP>2.0.ZU;2-5
Abstract
We report on results of tests of the 16-channel Hamamatsu R5900-M16 ph otomultiplier tube for gain uniformity and stability, pixel uniformity , linearity, cross-talk, and quantum efficiency. These parameters are essential to our application; i.e., the read out of signals which are generated by the passage of ionizing particles through the scintillati ng strips of an electromagnetic shower maximum detector and which are carried to the photodetector by optical fibers. In all respects the R5 900-M16 tube performance is found to be superior to that of earlier ge nerations of multianode photomultiplier tubes.