THE RESPONSE OF THE EUROBALL CLUSTER DETECTOR TO GAMMA-RADIATION UP TO 10 MEV

Citation
M. Wilhelm et al., THE RESPONSE OF THE EUROBALL CLUSTER DETECTOR TO GAMMA-RADIATION UP TO 10 MEV, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 381(2-3), 1996, pp. 462-465
Citations number
26
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
381
Issue
2-3
Year of publication
1996
Pages
462 - 465
Database
ISI
SICI code
0168-9002(1996)381:2-3<462:TROTEC>2.0.ZU;2-P
Abstract
The Euroball Cluster detector consists of seven closely packed hexagon al encapsulated germanium detectors which are mounted in a common cryo stat and surrounded by a EGO suppression shield, By adding back the Co mpton scattered events between the seven germanium detectors the detec tor system behaves similar as a single germanium detector with a volum e of 2000 cm(3). The response of this very large germanium detector to gamma-rays up to 10 MeV has been measured using a short lived Al-24 s ource produced at the Cologne FN Tandem accelerator. At 10 MeV the eff iciency of the Cluster detector is nearly one order of magnitude large r than that of a 30% germanium detector.