INFLUENCE OF THE THICKNESS AND THE ATTENUATION COEFFICIENT OF A BACKING ON THE RESPONSE OF TRANSDUCERS

Citation
J. Assaad et al., INFLUENCE OF THE THICKNESS AND THE ATTENUATION COEFFICIENT OF A BACKING ON THE RESPONSE OF TRANSDUCERS, Ultrasonics, 34(2-5), 1996, pp. 103-106
Citations number
4
Categorie Soggetti
Acoustics,"Radiology,Nuclear Medicine & Medical Imaging
Journal title
ISSN journal
0041624X
Volume
34
Issue
2-5
Year of publication
1996
Pages
103 - 106
Database
ISI
SICI code
0041-624X(1996)34:2-5<103:IOTTAT>2.0.ZU;2-L
Abstract
In this paper, the characteristic mechanical. impedance of an absorbin g medium (backing) is expressed in terms of its thickness T-B and the tangent of the phase angle (tan delta = m, delta is the loss angle). I n fact, the backing has been considered as a lossy transmission line. Then, using the model of Mason's circuit, the electrical impedance and radiated power of an in-air backed transducer will be expressed in te rms of tan delta and T-B. It is shown that the characteristics of such transducers do not change if the thickness is scaled down and the pro duct mT(B) stays constant. The same properties have been obtained for backed transducer bars using a finite element method (FEM) analysis.