Jc. Mohan et M. Bhargawa, DOPPLER-ECHOCARDIOGRAPHIC ASSESSMENT OF PROSTHETIC AORTIC-VALVE AREA - ESTIMATION WITH THE CONTINUITY EQUATION COMPARED TO THE GORLIN FORMULA, International journal of cardiology, 55(2), 1996, pp. 177-181
Effective orifice area of 3 different designs of prosthetic valves imp
lanted in the aortic position was determined by the continuity equatio
n and the Gorlin formula using Doppler hemodynamic data. The orifice a
rea by the two methods correlated well in the case of tilting disc pro
stheses (r = 0.75, P = 0.0001, n = 37, SEE = 0.17 cm(2)) but poorly in
the case of bileaflet mechanical valves (r = 0.40, P = 0.17, n = 13)
and ball-in-cage prostheses (r = 0.58, P = 0.06, n = 11). Estimation o
f prosthetic aortic valve area by the Gorlin formula is inappropriate
in the latter two types of prostheses because of design-related variab
le empiric constant.