C. Valiere et al., STABILITY AND COMPATIBILITY STUDY OF A CARBOPLATIN SOLUTION IN SYRINGES FOR CONTINUOUS AMBULATORY INFUSION, International journal of pharmaceutics, 138(1), 1996, pp. 125-128
The aim of this study was to verify the physico-chemical stability and
compatibility between a 10 mg/ml carboplatin parenteral solution and
the portable pump Intelliject(R) when stored at 25 degrees C over an 8
-day period. During this study, the pH was determined and carboplatin
degradation products were assayed by a validated HPLC method. The comp
atibility study consisted of the titration of reducing substances, TLC
of extracted additives, assay of extracted ions by AAS, resistivity m
easure, dynamometric test and UV, visible and IR absorption spectropho
tometry. Stability and compatibility studies showed that a 10 mg/ml ca
rboplatin parenteral solution remained chemically stable in the Intell
iject(R) pump over a period of 8 days under the tested conditions.