Tr. Gentile et al., REALIZATION OF A SCALE OF ABSOLUTE SPECTRAL RESPONSE USING THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY HIGH-ACCURACY CRYOGENIC RADIOMETER, Applied optics, 35(22), 1996, pp. 4392-4403
Using the National Institute of Standards and Technology high-accuracy
cryogenic radiometer (HACR), we have realized a scale of absolute spe
ctral response between 406 and 920 nm. The HACR, an electrical-substit
ution radiometer operating at cryogenic temperatures, achieves a combi
ned relative standard uncertainty of 0.021%. Silicon photodiode light-
trapping detectors were calibrated against the HACR with a typical rel
ative standard uncertainty of 0.03% at nine laser wavelengths between
406 and 920 nm. Modeling of the quantum efficiency of these detectors
yields their responsivity throughout this range with comparable accura
cy.