R. Perezolmos et al., DETERMINATION OF BORIC-ACID IN NICKEL-PLATING BATHS USING A TETRAFLUOROBORATE SELECTIVE ELECTRODE, Analusis, 24(4), 1996, pp. 143-145
Since the analytical control of boric acid contents in nickel plating
baths is important, we developed a potentiometric method, using tetraf
luoroborate selective electrodes. The boric acid contained in the samp
les was converted into tetrafluoroborate anions by means of a fluorifi
cation process; afterwards this ion was determined by the sample addit
ion method. The precision and accuracy of the recommended procedure, e
xpressed in terms of average variation coefficient (1.2%) and average
percentage of spike recovery (99.6%), were better than those obtained
by application, on the same samples, of the mannitol method (the refer
ence technique).