A flow system based on the monosegmented flow analysis (MSFA) approach
is described for delivery of liquid samples to arc/spark excitation s
ources commonly used in spectrographs. A Carl Zeiss PGS-2 spectrograph
, previously-automated in the laboratory by replacing its photographic
plate detection system with a photodiode array, was employed, The sam
ple is introduced via an injection port into the path to the excitatio
n source, where the liquid sample plug (typically 50 mu l) is passed t
hrough a hole drilled in a conventional graphite electrode, The arc/sp
ark is applied to excite the sample for a computer-controlled time int
erval and the light emitted is integrated, After sample measurement, t
he electrode is cleaned by applying the arc/spark while the mater stre
am, that follow the sample plug, is being pumped, About 25 samples can
be analysed per hour, The flow system was evaluated for determination
of aluminium (3-50 mg l(-1)), lead (5-160 mg l(-1)) and mixtures of h
afnium (60-120 mg l(-1)) and zirconium (600-1300 mg l(-1)).