AN INJECTION OLFACTOMETER FOR HUMANS AND A NEW METHOD FOR THE MEASUREMENT OF THE SHAPE OF THE OLFACTORY PULSE

Citation
Ra. Dewijk et al., AN INJECTION OLFACTOMETER FOR HUMANS AND A NEW METHOD FOR THE MEASUREMENT OF THE SHAPE OF THE OLFACTORY PULSE, Behavior research methods, instruments, & computers, 28(3), 1996, pp. 383-391
Citations number
17
Categorie Soggetti
Psychology, Experimental","Psychologym Experimental
ISSN journal
07433808
Volume
28
Issue
3
Year of publication
1996
Pages
383 - 391
Database
ISI
SICI code
0743-3808(1996)28:3<383:AIOFHA>2.0.ZU;2-K
Abstract
An injection olfactometer for human psychophysical experiments is desc ribed. The olfactometer is capable of presenting, either mono- or birh inally, single odorants and binary mixtures. The pulse form of the olf actory stimulus at the exit of the olfactometer approaches a rectangul ar form, which makes the olfactometer well suited for olfactory reacti on time measurements. Since stimulation periods are independent of the subject's inhalation, time-intensity measurements over prolonged peri ods of time are possible as well. In the second part of this paper, a new technique is described for the measurement of the stimulus pulse f orm at the exit of the olfactometer.