I. Wolman et al., TRANSVAGINAL ULTRASONOGRAPHIC MEASUREMENTS OF ENDOMETRIAL THICKNESS -A REPRODUCIBILITY STUDY, Journal of clinical ultrasound, 24(7), 1996, pp. 351-354
Citations number
12
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging",Acoustics
Objective: The study was undertaken to assess the reproducibility of e
ndometrial thickness measurements by transvaginal ultrasonography (TVS
). Methods: In a prospective blind study, two examiners measured the e
ndometrial thickness of 25 patients by TVS on two separate occasions 3
0 minutes apart. Results: The reliability test performed for each exam
iner was statistically less significant for the intra-observer variati
on of each observer (r = 0.95 and r = 0.93), than between both examine
rs (r = 0.85). Although there was no statistically significant differe
nce between the observations, the mean range of observations was 2.12
+ 1.27 mm. Conclusions: A safety margin of error should be taken into
consideration while recommending a cutoff under which no curettage is
needed. (C) 1996 John Wiley & Sons, Inc.