SIMPLIFICATION OF THE PROCEDURE FOR REUSE OF LIF-MG,CU,P TL DETECTORS

Citation
S. Wang et al., SIMPLIFICATION OF THE PROCEDURE FOR REUSE OF LIF-MG,CU,P TL DETECTORS, Radiation protection dosimetry, 65(1-4), 1996, pp. 211-212
Citations number
6
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
ISSN journal
01448420
Volume
65
Issue
1-4
Year of publication
1996
Part
1
Pages
211 - 212
Database
ISI
SICI code
0144-8420(1996)65:1-4<211:SOTPFR>2.0.ZU;2-X
Abstract
LiF(Mg,Cu,P) TL detectors have been re-used in two ways: without annea ling and with an annealing of 90 s in a TL reader in the dose range of 3 x 10(-4) - 0.1 Gy. No change of TL sensitivity and dosimetric prope rties was observed. The residual TL signal without annealing was about 4% after each measurement, and the coefficient of variation of the se nsitivity was 2.2% after ten cycles of re-use. The residual signal wit h TL reader annealing is between 0.16 and 0.23%. The simplified re-use procedure is of significance if it is applied properly to the routine applications of personal and environmental monitoring.