Gg. Cai et al., DOSE-RESPONSE OF THE INDIVIDUAL PEAK FOR LIF-MG,TI (TLD-100) AND LIF-MG,CU,P (GR-200A) TL MATERIALS AT THE PHOTON ENERGIES OF 30, 104, 1250KEV, Radiation protection dosimetry, 65(1-4), 1996, pp. 213-215
Citations number
10
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging","Nuclear Sciences & Tecnology
The dose responses for LiF:Mg,Ti (TLD-100) and LiF:Mg,Cu,P (GR-200A) h
ave been examined at the photon energies of 30, 104 and 1250 keV. A GC
A program has been used for the deconvolution of the glow peaks and th
e behaviours of the individual peaks have been analysed. For TLD-100,
an increasing supralinearity was observed for the higher temperature p
eaks, and the supralinearity at 30 keV was lower than that at 1250 keV
. For GR-200A, peak 4 showed supralinearity around 10 Gy. A possible c
orrelation seems to exist between the energy response and dose respons
e. A brief discussion and explanation are provided.