This report describes the case of a patient in whom atrial perforation
with penetration of the thoracic wall was diagnosed 2 years after the
implantation of an Accufix lead. Despite this complication, atrial de
tection in the bipolar mode and ventricular pacing were normal. Digita
l fluoroscopy detected a fracture with extrusion of a short segment of
the retention wire. The rupture of the retention wire might have been
the result, but was not the cause of the perforation.