V. Thompson et al., COMPARISON OF PORTAL IMAGING AND MEGAVOLTAGE VERIFICATION FILMS FOR CONFORMAL PELVIC IRRADIATION, British journal of radiology, 69(828), 1996, pp. 1191-1195
This study has evaluated the use of electronic portal imaging (EPI) as
a replacement for megavoltage verification films in the treatment of
conformal pelvic patients. The two methods of imaging have been compar
ed in 40 patients. Several advantages of EPI were identified, includin
g ease of image acquisition, regular assessment of reproducibility of
set-up and the possibility of on-line intervention when necessary. The
findings suggest that EPI is a suitable replacement for films.