K. Fujiwara et al., DETERMINATION OF SILICATE BY HOLLOW-CATHODE GLOW DISCHARGE-ATOMIC EMISSION-SPECTROMETRY WITH HYDRIDE GENERATION TECHNIQUE, Analytical letters, 29(11), 1996, pp. 1985-1992
Hollow cathode glow discharge atomic emission spectrometry has been ap
plied to the determination of silicon coupled with a novel gaseous hyd
ride generation technique, involving drying of an aqueous solution of
silicate (sample) and mixing with powdered LiAlH4. Sample introduction
into the glow discharge chamber was performed via a pinhole at the ce
nter of the cathode which was connected to the hydride generator. The
detection limit for silicon was 6 mu g at 288.1 nm and 30 mu g at 251.
6 nm.