DETERMINATION OF RETENTION INDEXES OF POLYCHLOROBIPHENYLS BY USING OTHER COMPOUNDS DETECTABLE BY ELECTRON-CAPTURE DETECTION OR SELECTED POLYCHLOROBIPHENYLS AS THE REFERENCE SERIES
G. Castello et G. Testini, DETERMINATION OF RETENTION INDEXES OF POLYCHLOROBIPHENYLS BY USING OTHER COMPOUNDS DETECTABLE BY ELECTRON-CAPTURE DETECTION OR SELECTED POLYCHLOROBIPHENYLS AS THE REFERENCE SERIES, Journal of chromatography, 741(2), 1996, pp. 241-249
Citations number
49
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
The highest sensitivity to polychlorobiphenyls (PCBs) is offered by el
ectron-capture detection (ECD), but the easy identification of individ
ual congeners remains unresolved. Published retention data are of limi
ted usefulness, as the lack of a uniform style and the dependence on t
he analysis conditions do not permit their general use. Retention inde
x values of PCBs cannot be easily referred to n-alkanes, due to the ve
ry low response of these compounds to ECD, whereas other well detected
reference series can be used: n-alkyl iodides were found useful for t
he determination of retention indices of the first congeners of the PC
B series. Retention indices obtained by using selected DCMA (Dry Colou
r Manufacturers' Association) PCB reference mixtures can be directly u
sed or converted to standard Kovats indices. A mixture of PCBs that sh
ows a linear dependence of retention on the number of chlorine atoms o
n both isothermal and programmed-temperature analysis is suggested for
reference purposes. Other methods and standard mixtures used for iden
tification and quantitation purposes are reviewed and discussed.