We have used noise-equivalent count (NEC) rates to optimize count rate
performance for 3D acquisition in PET in a wide range of situations,
with particular reference to imaging of the torso. We have also compar
ed NEC performance for 2D and 3D acquisition in order to establish the
conditions under which 3D mode offers an improvement over 2D mode. Me
asurements were performed on four tissue-equivalent phantoms ranging i
n size from that of an infant's head (13 cm diameter) to that of an ob
ese adult's chest (37 cm x 48 cm). Count rate data were acquired as a
function of phantom size, activity in the field of view, lower energy
discriminator level (LLD) and acquisition mode, and NEC rates were der
ived as a function of these variables. The LLD at which the highest NE
C rate is obtained shows a dependence both on phantom size and on the
activity in the field of view both for 2D and for 3D acquisition. The
relative advantage of 3D mode over 2D mode, at the optimum LLD setting
, is also strongly dependent both on activity in the field of view (FO
V) and on the phantom size. The main limiting factors for 3D NEC rates
are detector dead-time for small phantoms and random coincidences for
large phantoms. The 3D NEC rate is more than twice as great as the 2D
NEC rate when less than 60 MBq is present in the FOV for all phantoms
except the largest, in which case a ratio of two is only achieved for
activities less than 25 MBq. For the smallest phantom, 3D/2D NEC rati
os of greater than 3.5 are obtained when the activity in the FOV falls
below 10 MBq.