AN AUTOMATED SAMPLE PREPARATION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF MK-591, A NOVEL LEUKOTRIENE BIOSYNTHESIS INHIBITOR, IN HUMAN PLASMA

Citation
C. Lin et al., AN AUTOMATED SAMPLE PREPARATION AND HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF MK-591, A NOVEL LEUKOTRIENE BIOSYNTHESIS INHIBITOR, IN HUMAN PLASMA, Journal of pharmaceutical and biomedical analysis, 14(11), 1996, pp. 1601-1608
Citations number
11
Categorie Soggetti
Pharmacology & Pharmacy
ISSN journal
07317085
Volume
14
Issue
11
Year of publication
1996
Pages
1601 - 1608
Database
ISI
SICI code
0731-7085(1996)14:11<1601:AASPAH>2.0.ZU;2-2
Abstract
A high performance liquid chromatography assay utilizing an automated sample preparation procedure for the determination of a novel leukotri ene biosynthesis inhibitor, (MK-591), in human plasma has been develop ed. After aliquoting plasma samples and adding internal standard manua lly, the BenchMate (TM) Workstation executed protein precipitation and solid-phase extraction. Following evaporation to dryness, the residue was reconstituted and chromatographed isocratically on a cyano-phase analytical column. MK-591 and the internal standard were separated fro m each other and from endogenous plasma substances and detected with a n absorbance detector. The assay has been validated in the concentrati on range 10-1000 ng ml(-1) and has the sensitivity and specificity nec essary to quantify plasma concentrations from several clinical studies .