Ka. Ellenbogen et al., ACUTE EFFECTS OF RADIOFREQUENCY ABLATION OF ATRIAL ARRHYTHMIAS ON IMPLANTED PERMANENT PACING SYSTEMS, PACE, 19(9), 1996, pp. 1287-1295
We studied the safety of performing RF catheter ablation in patients w
ith implanted permanent pacemakers by monitoring the function of impla
nted pacing systems before, during, and immediately after exposure to
RF energy. Patients with implanted pacing systems may require RF ablat
ion for treatment of a variety of tachyarrhythmias, High frequency ele
ctromagnetic fields, such as RF energy, may affect implanted pacing sy
stems, causing temporary or permanent loss of output, undersensing, ov
ersensing, asynchronous pacing, or reversion to ''reset'' (Recommended
Replacement Time or Power On Reset) parameters. Thirty-five patients
with implanted pacing systems (23 DDDR, 6 VVIR, 5 DDD, 1 VVI, 31 bipol
ar and 4 unipolar) underwent RF catheter ablation. Prior to ablation,
each pacing system underwent measurements of pacing and sensing thresh
olds, telemetry of intracardiac electrograms and measurement of batter
y voltage and lead impedance(s). During ablation, pacemaker function w
as monitored by real-time telemetry, intracardiac electrograms, and su
rface EGG. Immediately after ablation, each pacing system was reevalua
ted. Telemetry during RF ablation revealed normal pacing and sensing i
n 14 (40%) of 35 patients. Refractory period extension with asynchrono
us pacing and noise mode reversion were seen in 16 (46%) of 35 patient
s. Rare under- and/or oversensing, reversion to reset parameters, and
telemetry ''lock up'' with inhibition of pacing output was seen in a f
ew patients. After ablation, there were no significant changes in atri
al or ventricular pacing or sensing thresholds or measurements of atri
al and ventricular lead impedances. We conclude that most permanent pa
cemakers are not adversely affected by exposure to RF energy during ca
theter ablation. A variety of pacemaker behaviors may be seen during R
F ablation, and a thorough understanding of each pulse generator's pot
ential response(s) to electromagnetic interference is important before
undertaking catheter ablation in patients with permanent pacemakers.
Careful reevaluation of the patient's pacing system following the proc
edure is mandatory.