USE OF YB-169 AS A CALIBRATION STANDARD

Citation
S. Bhattacharya et al., USE OF YB-169 AS A CALIBRATION STANDARD, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 378(3), 1996, pp. 515-517
Citations number
10
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
378
Issue
3
Year of publication
1996
Pages
515 - 517
Database
ISI
SICI code
0168-9002(1996)378:3<515:UOYAAC>2.0.ZU;2-G
Abstract
The energies and intensities of the gamma and X-rays in the decay of Y b-169 (32d) have been measured with a low energy photon spectrometer ( LEPS) and a high purity germanium (HPGe) detector. This source is reco mmended for use as a standard for low-energy efficiency calibration of such detectors. It is also proposed to use this source as a standard for conversion electrons, as the multipolarities and mixing ratios are well known for most strong transitions.