T. Lakatos et al., A SIMPLE PULSED DRAIN FEEDBACK PREAMPLIFIER FOR HIGH-RESOLUTION HIGH-RATE NUCLEAR SPECTROSCOPY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 378(3), 1996, pp. 583-588
A non-optical, pulsed reset technique for use in charge sensitive prea
mplifiers with Si(Li) X-ray semiconductor detectors of conventional po
larity is described. The phenomenon of impact ionization in the field
effect transistor is used for charge restoration. The process is contr
olled via the detector capacitance. The method needs no extra componen
ts in the input circuitry. No significant resolution degradation can b
e detected up to 600 kcps Mn K intensity if following the 5 mu s reset
ting pulse the signal processor is inhibited for 50 mu s.