G. Bertuccio et al., PERFORMANCE OF SI LEC GAAS DETECTORS AT 20-DEGREES-C AND -30-DEGREES-C FOR X-RAY AND GAMMA-RAY SPECTROSCOPY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 379(1), 1996, pp. 152-154
We report on the performance and limitations of a 7 mm(2), 100 mu m th
ick SI LEC GaAs detector, operated in overdepleted conditions with app
lied bias of 400 V at 20 degrees C and -30 degrees C, In detecting X-
and gamma-rays. At 20 degrees C performance is limited to a charge col
lection efficiency of 92% and to an energy resolution of about 16.5 ke
V FWHM on the 59.5 keV line of Am-241. At -30 degrees C the charge col
lection efficiency is 85% and a resolution in between 1.8 and 2.2 keV
FWHM in the photon energy range 8-60 keV has been achieved. The latest
data represent to bur knowledge the best ones yet reported for GaAs L
EC detectors.