EXTERNAL FIXATION FOR COMMINUTED PHALANGEAL FRACTURES - A BIOMECHANICAL CADAVER STUDY

Citation
F. Fitoussi et al., EXTERNAL FIXATION FOR COMMINUTED PHALANGEAL FRACTURES - A BIOMECHANICAL CADAVER STUDY, Journal of hand surgery. British volume, 21B(6), 1996, pp. 760-764
Citations number
22
Categorie Soggetti
Orthopedics,Surgery
ISSN journal
02667681
Volume
21B
Issue
6
Year of publication
1996
Pages
760 - 764
Database
ISI
SICI code
0266-7681(1996)21B:6<760:EFFCPF>2.0.ZU;2-T
Abstract
The mechanical rigidity obtained by external fixation in a comminuted phalangeal fracture model was assessed and the results compared with t wo other types of internal fixation commonly used (lateral plate and c rossed Kirschner wires) in a biomechanical cadaver study. Each fixatio n technique was tested in apex palmar bending, compression and torsion , The results showed that lateral plating provided the best rigidity i n apex palmar bending and compression and that external fixation and K irschner wires showed the same mechanical properties, For the torque t est, external fixation provided the best rigidity.