R. Thottappillil et al., PROPERTIES OF M-COMPONENTS FROM CURRENTS MEASURED AT TRIGGERED LIGHTNING CHANNEL BASE, JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 100(D12), 1995, pp. 25711-25720
Channel base currents from triggered lightning were measured at the NA
SA Kennedy Space Center, Florida, during summer 1990 and at Fort McCle
llan, Alabama, during summer 1991. An analysis of the return stroke da
ta and overall continuing current data has been published by Fisher et
al. [1993]. Here an analysis is given of the impulsive processes, cal
led M components, that occur during the continuing current following r
eturn strokes. The 14 flashes analyzed contain 37 leader-return stroke
sequences and 158 M components, both processes lowering negative char
ge from cloud to ground. Statistics are presented for the following M
current pulse parameters: magnitude, rise time, duration, half-peak wi
dth, preceding continuing current level, M interval, elapsed time sinc
e the return stroke, and charge transferred by the M current pulse. A
typical M component in triggered lightning is characterized by a more
or less symmetrical current pulse having an amplitude of 100-200 A (2
orders of magnitude lower than that for a typical return stroke [Fishe
r et al., 1993]), a 10-90% rise time of 300-500 mu s (3 orders of magn
itude larger than that for a typical return stroke [Fisher et al., 199
3]), and a charge transfer to ground of the order of 0.1 to 0.2 C (1 o
rder of magnitude smaller than that for a typical subsequent return st
roke pulse [Berger et al., 1975]). About one third of M components tra
nsferred charge greater than the minimum charge reported by Berger et
al. [1975] for subsequent leader-return stroke sequences. No correlati
on was found between either the M charge or the magnitude of the M com
ponent current (the two are moderately correlated) and any other param
eter considered. M current pulses occurring soon after the return stro
ke tend to have shorter rise times, shorter durations, and shorter M i
ntervals than those which occur later. M current pulses were observed
to be superimposed on continuing currents greater than 30 A or so, wit
h one exception out of 140 cases, wherein the continuing current level
was measured to be about 20 A. The first M component virtually always
(one exception out of 34 cases) occurred within 4 ms of the return st
roke. This relatively short separation time between return stroke and
the first M component, coupled with the observation of Fisher et al. [
1993] that continuing currents lasting longer than 10 ms never occur w
ithout M current pulses, implies that the M component is a necessary f
eature of the continuing current mode of charge transfer to ground.