DETERMINATION OF ACETYLCHOLINE BY ONLINE MICRODIALYSIS COUPLED WITH PRE-MICROBORE AND POST-MICROBORE COLUMN ENZYME REACTORS WITH ELECTROCHEMICAL DETECTION

Citation
Tr. Tsai et al., DETERMINATION OF ACETYLCHOLINE BY ONLINE MICRODIALYSIS COUPLED WITH PRE-MICROBORE AND POST-MICROBORE COLUMN ENZYME REACTORS WITH ELECTROCHEMICAL DETECTION, Journal of chromatography B. Biomedical applications, 678(2), 1996, pp. 151-155
Citations number
15
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Journal of chromatography B. Biomedical applications
ISSN journal
15726495 → ACNP
Volume
678
Issue
2
Year of publication
1996
Pages
151 - 155
Database
ISI
SICI code
Abstract
A sensitive procedure consisting of a pre- and post-microbore column r eactor sequence of a LC-electrochemical detection system coupled with on-line microdialysis system is described in the present study to meas ure endogenous acetylcholine concentration in freely moving rats. The pre-column packed, with immobilized choline oxidase and catalase, was used to remove choline, whereas the post-column, packed with immobiliz ed acetylcholine oxidase and choline oxidase, was used to measure acet ylcholine selectively, The detection limit of acetylcholine was found to be 5 fmol/mu l (50 fmol/10 mu l). The usefulness of the described m ethodology was evaluated by examining the change in the striatal acety lcholine concentration of freely moving rats after physostigmine (0.5 mg/kg, s.c.) administration.