HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF MUPIROCIN IN POLYETHYLENE GLYCOL-400 AND GLYCOL-3350 USING DUAL ULTRAVIOLET AND EVAPORATIVE LIGHT-SCATTERING DETECTION

Authors
Citation
Rs. Porter et Tk. Chen, HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF MUPIROCIN IN POLYETHYLENE GLYCOL-400 AND GLYCOL-3350 USING DUAL ULTRAVIOLET AND EVAPORATIVE LIGHT-SCATTERING DETECTION, Journal of chromatography, 732(2), 1996, pp. 399-402
Citations number
6
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Volume
732
Issue
2
Year of publication
1996
Pages
399 - 402
Database
ISI
SICI code
Abstract
Using dual detection with UV and evaporative light scattering detector s, a gradient LC-MS-compatible method was successfully developed for t he separation and identification of impurity and degradates in mupiroc in in a complex matrix containing 98% (w/w) of polyethylene glycol (PE G) 400 and PEG 3350. The method enabled the eluent containing the PEGs to be diverted before it entered and contaminated the mass spectromet er source. The impurity was identified as Pseudomonic D, and the degra dates were rearrangement product I and rearrangement product II, respe ctively.