COMPARISONS OF SOME NIST FIXED-POINT CELLS WITH SIMILAR CELLS OF OTHER STANDARDS LABORATORIES

Citation
Bw. Mangum et al., COMPARISONS OF SOME NIST FIXED-POINT CELLS WITH SIMILAR CELLS OF OTHER STANDARDS LABORATORIES, Metrologia, 33(3), 1996, pp. 215-225
Citations number
22
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
Journal title
ISSN journal
00261394
Volume
33
Issue
3
Year of publication
1996
Pages
215 - 225
Database
ISI
SICI code
0026-1394(1996)33:3<215:COSNFC>2.0.ZU;2-T
Abstract
In this paper we present results of international comparisons of fixed -point cells of some of the defining fixed-point materials of the Inte rnational Temperature Scale of 1990, These comparisons involved cells from seven national laboratories, although in some cases only one type of fixed-point material was compared, Except for silver cells, the ag reement among cells of the same defining fixed-point material from the various laboratories was to within 1 mK. The expanded uncertainties ( k = 2) of the comparison measurements were 12 mu K for Ga, 18 mu K for H2O, 27 mu K for Sn, 35 mu K for In, 37 mu K for Zn, 42 mu K for Al, and 55 mu K for Ag.