COMPARISON OF THE BIPM CRYOGENIC RADIOMETER WITH A MECHANICALLY COOLED CRYOGENIC RADIOMETER FROM THE NPL

Citation
R. Goebel et al., COMPARISON OF THE BIPM CRYOGENIC RADIOMETER WITH A MECHANICALLY COOLED CRYOGENIC RADIOMETER FROM THE NPL, Metrologia, 33(2), 1996, pp. 177-179
Citations number
4
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
Journal title
ISSN journal
00261394
Volume
33
Issue
2
Year of publication
1996
Pages
177 - 179
Database
ISI
SICI code
0026-1394(1996)33:2<177:COTBCR>2.0.ZU;2-4
Abstract
The direct comparison at the Bureau International des Poids et Mesures (BIPM) of the BIPM cryogenic radiometer (Radiox) with a mechanically cooled cryogenic radiometer (MCR) from the National Physical Laborator y (NPL) is described. The radiometers were compared using a laser beam at lambda = 647 nm with power levels of 400 mu W and 650 mu W. They w ere found to be in good agreement, as the measured difference in respo nsivity was found to be 1,5 parts in 10(5), with a combined standard u ncertainty of 8 parts in 10(5).