L. Yang et Pt. Kissinger, DETERMINATION OF OXIDASE ENZYME SUBSTRATES USING CROSS-FLOW IN THIN-LAYER AMPEROMETRY, Electroanalysis, 8(8-9), 1996, pp. 716-721
Thin-layer cross-flow amperometric detectors with enzyme modified work
ing electrodes were developed for the determination of oxidase substra
tes in flowing streams. The enzyme was covalently immobilized in an os
mium redux polymer film on the electrode surface. Electrodes modified
with glucose oxidase or lactate oxidase were prepared for the continuo
us monitoring of glucose or lactate. A peroxidase electrode coupled wi
th oxidase enzyme reactors was also explored for the determination of
glucose, lactate, and acetylcholine/choline in continuous monitoring s
ystems or following liquid chromatography. The effects of interference
s, oxygen, mass transfer limiting membrane, and perfusate composition
on electrode performance an discussed.