F. Mizutani et al., FLOW-INJECTION ANALYSIS OF L-LACTIC ACID USING AN ENZYME-POLYION COMPLEX-COATED ELECTRODE AS THE DETECTOR, Talanta, 43(10), 1996, pp. 1815-1820
The concentration of L-lactic acid was determined by a combination of
flow injection analysis with amperometric enzyme sensor detection. The
enzyme sensor was prepared by immobilizing lactate oxidase in a layer
of polyion complex consisting of poly-L-lysine and poly(4-styrenesulf
onate). The sensor-based system can be used for the determination of L
-lactate concentration up to 6 mM with a sampling rate of 120 h(-1), a
nd is stable for 8 weeks after 1000 L-lactate injections. The permsele
ctivity of the polyion complex matrix is effective for reducing the re
sponse from electrochemical interferents such as L-ascorbic acid, uric
acid and acetaminophen.