Cc. Vera et al., A 3-POINT CALIBRATION PROCEDURE FOR MATRIX-ASSISTED LASER DESORPTION IONIZATION MASS-SPECTROMETRY UTILIZING MULTIPLY-CHARGED IONS AND THEIRMEAN INITIAL VELOCITIES/, Rapid communications in mass spectrometry, 10(12), 1996, pp. 1429-1432
This paper reports the results obtained by using a new calibration pro
cedure for linear time-of-flight matrix-assisted laser desorption/ioni
zation mass spectrometry (MALDI). The procedure takes into account the
usual MALDI spectra structure by considering the multiply charged mol
ecular ions and the mean initial velocities as a self calibrating prop
erty. The mean velocities for all molecular ions are taken to be the s
ame and the procedure is based on the solution of a number of equation
s (at least three) which relate time-of-flight to mass. Three times be
tter accuracy was obtained in comparison with the normal internal cali
bration procedure.