SENSITIVITY-ENHANCED ECHO-PLANAR MRI AT 1.5T USING A 5X5-MESH DOME RESONATOR

Citation
Kl. Meyer et al., SENSITIVITY-ENHANCED ECHO-PLANAR MRI AT 1.5T USING A 5X5-MESH DOME RESONATOR, Magnetic resonance in medicine, 36(4), 1996, pp. 606-612
Citations number
18
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
07403194
Volume
36
Issue
4
Year of publication
1996
Pages
606 - 612
Database
ISI
SICI code
0740-3194(1996)36:4<606:SEMA1U>2.0.ZU;2-X
Abstract
In this work a 5 x 5 mesh dome resonator that has been optimized for f unctional brain imaging is presented, The resonator was reduced in len gth and diameter compared with previous versions to reduce sample loss es, thus enhancing the signal-to-noise ratio of the acquired data, In addition, a 5 x 5 mesh design was employed, which offered improved axi al homogeneity over an earlier 3 x 3 mesh version, The new resonator e xhibited high sensitivity and good homogeneity over the brain volume, permitting analysis of functional activation over large areas of the c erebral cortex. In a direct comparison with a standard clinical head-i maging resonator, the high sensitivity of the 5 x 5 mesh dome resonato r resulted in greater statistical confidence in functional activation.