MONITORING OF DISSOLVED REACTIVE PHOSPHORUS IN WASTEWATERS BY FLOW-INJECTION ANALYSIS .2. ONLINE MONITORING-SYSTEM

Citation
Rl. Benson et al., MONITORING OF DISSOLVED REACTIVE PHOSPHORUS IN WASTEWATERS BY FLOW-INJECTION ANALYSIS .2. ONLINE MONITORING-SYSTEM, Water research, 30(9), 1996, pp. 1965-1971
Citations number
7
Categorie Soggetti
Engineering, Civil","Environmental Sciences","Water Resources
Journal title
ISSN journal
00431354
Volume
30
Issue
9
Year of publication
1996
Pages
1965 - 1971
Database
ISI
SICI code
0043-1354(1996)30:9<1965:MODRPI>2.0.ZU;2-E
Abstract
An automated-flow injection (FI) analyser for the determination of dis solved reactive phosphorus (DRP) in wastewaters is described. The anal ytical method is based on the formation of molybdenum blue. The limits of detection using a solid-state photometric detector (LED-photodiode ) were 0.05 mg P 1(-1) and the linear range was 0-20 mg P 1(-1). Detai ls of construction, field-testing (including sample filtration) and va lidation are given. The analyser proved very reliable during an unatte nded operational period of seven days. Operational costs were minimise d, with low reagent consumption and minimal operator intervention. Dat a retrieval and analyser status were remotely interrogated by means of a cellular telephone-based communications link. The real-time results from the analyser showed good correlation with those generated by a v alidated laboratory method. Copyright (C) 1996 Elsevier Science Ltd