Dk. Montague et al., AMS 3-PIECE INFLATABLE PENILE PROSTHESIS IMPLANTATION IN MEN WITH PEYRONIES-DISEASE - COMPARISON OF CX AND ULTREX CYLINDERS, The Journal of urology, 156(5), 1996, pp. 1633-1635
Purpose: We compared the ability of the CX and Ultrex cylinders to str
aighten the penis in men who received a 3-piece AMS 700 series inflat
able penile prosthesis for erectile dysfunction and erectile deformity
due to Peyronie's disease. Materials and Methods: The records of 34 a
nd 38 patients receiving devices with CX and Ultrex cylinders, respect
ively, were reviewed. Results: All 34 patients receiving the CX cylind
ers achieved complete penile straightening with cylinder inflation and
bending alone. In 10 of 38 patients receiving the Ultrex cylinders co
mplete straightening with cylinder inflation and bending could not be
achieved, and simultaneous corporoplasty was necessary. Conclusions: G
irth expanding CX cylinders have better penile straightening propertie
s than girth and length expanding Ultrex cylinders. We recommend use o
f CX cylinders for 3-piece AMS 700 series inflatable penile prosthesis
implantation in men with Peyronie's disease.