DETERMINATION OF CYANIDE IN MICROSAMPLES BY MEANS OF CAPILLARY-FLOW INJECTION-ANALYSIS WITH AMPEROMETRIC DETECTION

Citation
U. Backofen et al., DETERMINATION OF CYANIDE IN MICROSAMPLES BY MEANS OF CAPILLARY-FLOW INJECTION-ANALYSIS WITH AMPEROMETRIC DETECTION, Fresenius' journal of analytical chemistry, 356(3-4), 1996, pp. 271-273
Citations number
11
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
356
Issue
3-4
Year of publication
1996
Pages
271 - 273
Database
ISI
SICI code
0937-0633(1996)356:3-4<271:DOCIMB>2.0.ZU;2-5
Abstract
A new approach for determining cyanide in microsamples is described. T he method is based on capillary flow injection analysis (CFIA) with am perometric detection. The sensing electrode is a silver-plated microdi sk electrode, where cyanide can react under formation of a dicyanoarge ntate complex. A remarkably low mass detection limit of 231 fmol cyani de is obtained for an injection volume of 60 nl. The sample throughput of the CFIA-arrangement is comparable with a conventional sized FIA-s ystem. A practical application is given by analyzing the cyanide (amyg dalin) concentration in apple kernels.