TEMPERATURE CONTROL AND THERMAL DOSIMETRY BY MICROWAVE RADIOMETRY IN HYPERTHERMIA

Citation
L. Dubois et al., TEMPERATURE CONTROL AND THERMAL DOSIMETRY BY MICROWAVE RADIOMETRY IN HYPERTHERMIA, IEEE transactions on microwave theory and techniques, 44(10), 1996, pp. 1755-1761
Citations number
16
Categorie Soggetti
Engineering, Eletrical & Electronic
ISSN journal
00189480
Volume
44
Issue
10
Year of publication
1996
Part
2
Pages
1755 - 1761
Database
ISI
SICI code
0018-9480(1996)44:10<1755:TCATDB>2.0.ZU;2-6
Abstract
This paper presents a synthesis of works undertaken by the Hyperthermi a Group of Lille (France) concerning the utilization of the microwave radiometry for the temperature control in hyperthermia therapy. This t echnique of noninvasive temperature control within the biological tiss ues has been integrated on many hyperthermia systems now commercialize d. We describe the principle of a new radiometer as well as the calcul ation of radiometric signals, They allow a noninvasive determination o f thermal maps inside tissues during the hyperthermia treatments. Many comparisons between theory and experiment have validated models of th ermal dosimetry whose provide a quantitative guidance for the planning of hyperthermia treatments.