K. Meier et al., BROAD-BAND CALIBRATION OF E-FIELD PROBES IN LOSSY MEDIA, IEEE transactions on microwave theory and techniques, 44(10), 1996, pp. 1954-1962
A broadband calibration procedure for E-field that minimizes the overa
ll uncertainties inherent in measurements in lossy dielectric liquids
has been developed. The analysis of the calibration requirements shows
that probes that are symmetrical with respect to their axis greatly f
acilitate accurate calibration, since the calibration procedure can be
divided into several discrete steps, Such a procedure is presented an
d analyzed with respect to its uncertainties. Absolute calibration is
performed at three frequency bands utilized in Europe for mobile commu
nications (450 Mhz. 900 MHz, and 1.8 GHz) and in different tissue-simu
lating liquids, The parameters obtained are verified by numerical simu
lations of the probe in the surrounding media, Such simulations allow
the assessment of some of the calibration parameters with sufficient a
ccuracy in cases where the experimental determination would be too ted
ious and time consuming.