HIGH-SENSITIVE DETECTION OF TRACE GASES USING OPTICAL HETERODYNE METHOD WITH A HIGH FINESSE INTRACAVITY RESONATOR

Citation
A. Mugino et al., HIGH-SENSITIVE DETECTION OF TRACE GASES USING OPTICAL HETERODYNE METHOD WITH A HIGH FINESSE INTRACAVITY RESONATOR, Optical review, 3(4), 1996, pp. 243-250
Citations number
21
Categorie Soggetti
Optics
Journal title
ISSN journal
13406000
Volume
3
Issue
4
Year of publication
1996
Pages
243 - 250
Database
ISI
SICI code
1340-6000(1996)3:4<243:HDOTGU>2.0.ZU;2-Y
Abstract
A novel highly sensitive method to detect the concentration of trace g as for environmental purposes is proposed. This method includes the ap plication of frequency modulation for stabilization of laser frequency , together with optical heterodyne measurement. The shift of resonance frequency of the Fabry-Perot cavity resonator is due to the change of refractive index, accompanied by the dispersion effect due to the pre sence of a target gas (e.g., methane, carbon dioxide or nitrogen dioxi de) inside the Fabry-Perot cavity.